TBB Global Việt Nam

TBB Global Việt Nam

TBB Global Việt Nam

TBB Global Việt Nam
TBB Global Việt Nam

Product detail

cs-2000
  • Views:
    779
Keyword tag: ,

ELTRA's CS-2000 is the only analyzer on the market for thedetermination of carbon and sulfur in organic as well as inorganic samples. For this purpose, the CS-2000 is equipped with both an induction and a resistance furnace (ELTRA Dual Furnace Technology), covering the full range of carbon and sulfur analysis.

The CS-2000 is available with up to four independent infrared cells, which allow the precise and simultaneous analysis of high and low carbon and/or sulfur concentrations. The sensitivity of the cells can be customized individually by selecting the length of the IR-paths to ensure the optimum measuring range for each application.

Application Examples

alloys, ashes, carbides, cast iron, cement, ceramics, coal, coke, copper, glass, gypsum, iron, limestone, metals, minerals, oil, ores, plant materials, refractory metals, rubber, sand, soils, steel, titanium, tobacco, ...

Product Advantages

  • full flexibility due to combined induction and resistance furnace: ELTRA Dual Furnace (EDF) Technology
  • analysis of organic and inorganic samples
  • due to ELTRA Dual Furnace (EDF) Technology: fractional analysis of carbon and sulfur
  • up to four independent infrared cells with flexible measuring ranges
  • due to gold IR path, increased cell live time for analysis of halogen or acid containig samples
  • automatic induction furnace cleaning
  • wide range of materials (organic and inorganic) can be analyzed
  • rapid, precise, accurate and reliable element determination
  • powerful (2.2 kW) induction furnace for temperatures above 2,000 °C
  • temperature of resistance furnace adjustable up to 1,550 °C in steps of 1 °C
  • effective, easily accessible dust trap
  • single and multipoint calibration
  • simultaneous calibration of carbon and sulfur
  • low maintenance
  • robust design allows usage in production control and laboratory
  • simultaneous carbon and sulfur determination with minimal sample preparation

Features

Measured elements carbon, sulfur
Samples inorganic, organic
Furnace alignment horizontal (resistance furnace) and vertical (induction furnace)
Sample carrier ceramic boats / crucibles
Field of application agriculture, chemistry / plastics, coal / power plant, construction materials, engineering / electronics, environment / recycling, geology / mining, glass / ceramics, medicine / pharmaceuticals, steel / metallurgy
Furnaces induction furnace, above 2,000 °C
resistance furnace (ceramic tube), adjustable up to 1,550 °C (steps of 1 °C)
Detection method solid state infrared absorption
Number of IR cells 1 - 4
Material of IR path gold
Typical analysis time induction furnace 40 - 50 s
resistance furnace 60 - 120 s
Chemicals required copper oxide, magnesium perchlorate, sodium hydroxide
Gas required compressed air (4 - 6 bar / 60 - 90 psi)
oxygen 99.5 % pure (2 - 4 bar / 30 - 60 psi)
Power requirements induction furnace 230 V, 50/60 Hz; max. 15 A, 3450 W
Power requirements resistance furnace 230 V, 50/60 Hz, max. heating up current 20 A
Dimensions (W x H x D) 88 x 80 x 60 cm
Weight ~ 150 kg
Required equipment balance (resolution 0.0001g), monitor, PC
Optional accessories Autoloader (for 36 or 130 crucibles), carrier gas purification, halogen trap, HTF-540 pre-heating furnace, voltage stabilizer 5 KVA

Function Principle

Operation CS-2000

In addition to the induction furnace (see CS-800) the CS-2000 also features a resistance furnace. The temperature can be selected in steps of 1 °C up to 1,550 C. The samples (such as coal) are weighed in ceramic boats. The weight is transferred from the interfaced balance to the PC or can be entered manually. The sample is then placed directly in the furnace for combustion, no accelerators are required. The typical analysis time is 60 to 180 seconds depending on the sample material. The combustion gases pass through the infrared cells for detection. There the signals are evaluated and the results are displayed automatically. Connection to a laboratory information management system (LIMS) is possible. The CS-2000 requires minimum maintenance and the chemicals which need to be maintained are easily accessible.

Measuring principle CS-2000

The measuring principle of the induction furnace corresponds to that of the CS-800. In the additional resistance furnace of the CS-2000 the sample is combusted in a pure oxygen stream, causing sulfur to react to sulfur dioxide (SO2), and carbon to react to carbon dioxide (CO2). Any moisture in the combustion gas is removed by magnesium perchlorate and the gas then passes into the detection unit. There the detection of the sulfur dioxide takes place in infrared cells. The CS-2000 allows for the combination of cells with different sensitivities (high/low) which makes it easily adaptable to the user’s requirements. After oxidation of sulfur dioxide to sulfur trioxide, SO3gas is removed with cellulose wool. Subsequently the carbon content is determined by infrared cells with customized measuring ranges.

Download
For information we will send you the latest information !
Copyright © 2018 Công Nghệ Toàn Cầu TBB . Designed by Nina.vn
Online : 2| This weed : 2668| Total : 986042
1
icon_zalod